ANAESTHETIC AND RESPIRATORY EQUIPMENT - CONICAL CONNECTORS - PART 1: CONES AND SOCKETS (ISO 5356-1:2004, IDT)
This Malaysian Standard specifies dimensional and gauging requirements for cones and sockets intended for connecting anaesthetic and respiratory equipment, e.g. in breathing systems, anaesthetic-gas scavenging systems and vaporizers.
This Malaysian Standard gives requirements for the following conical connectors:
a) 8.5 mm size intended for use in paediatric breathing systems;
b) 15 mm and 22 mm sizes intended for general use in breathing systems;
c) 22 mm latching connectors (including performance requirements);
d) 23 mm size intended for use with vaporizers, but not for use in breathing systems;
e) 30 mm size intended for the connection of a breathing system to an anaesthetic gas scavenging system.
This standard does not specify the medical devices and accessories on which these connections are to be provided.
Requirements for the application of conical connectors are not included in this standard, but are or will be given in the relevant standards for specific medical dev